Chapter 5 Calibration
SM25 User’s Manual
5.3.1.5 Step 5
Make sure the return value is close to 0V(tolerance < 1mV) when the system is stable. Then click “Yes”
button if it is true or click “No” button if not, see Fig 5-14.
Figure 5-14. Check Return Value.
5.3.1.6 Step 6
Make sure the return value is close to 20 V(tolerance < 1mV) when the system is stable. Then click “Yes”
button if it is true or click “No” button if not, see Fig 5-15.
Figure 5-15. Check 80% Full Scale Return Value.
Rev 1.0
5-10