When regulation reaches the edge of kV (40kV for the minimum, 100kV, 110kV, 120kV for the
maximum) the mA are changed (40kV mA decreased, Max kV mA increased)
5.2 Intermittent Fluoroscopy
When High Voltage Generator R306.3x or R306.4x is installed with a television chain, which uses
the Last image Hold feature, it is possible to configure a special automatism to reduce the patient
dose in case the exam is not dynamic.
Generally used in Bone Exams this technique is known also as Orthopedic Fluoroscopy;
Intermittent Fluoroscopy terms stands for a series of continuous emission intervals automatically
generated by the generator.
The interval emission is enabled after an initial period of image stabilization and it is suspended for
a continuous emission if a re-stabilization is need due to patient ROI change or movement.
During installation phase the duty cycle of the intermittent period is chosen in term of LIGHT
(values in the 0,1s ~ 1,5s interval) and DARK (values in the 0,1s ~ 1,5s interval).
Intermittent fluoroscopy can be both Chained parameters and Free parameters configurable.
5.3 High Contrast Fluoroscopy*
*High Contrast Fluoroscopy is an R306.4x option
When High Contrast Fluoroscopy is configured in current working station the relative icon
is shown in the top of the section.
Pressing the
key it is possible to access the High Contrast Fluoroscopy panel.
1
2
The panel, according to installation option can assume four different shapes:
High Contrast Fluoroscopy, Automatic (1) o Manual (2).
The presence of Automatic Brightness Control depends on the Fluoroscopy Television Controller
installed in the room, the choice between Manual and Automatic is made runtime by the user.
Note: The control gauges presents on the panel depends on the grabbing system connected to the
generator, more information are present in Digital System Drivers section of this manual.
0078 Ver ENG Rev K7
page 45