Download 1405-D - User Manual
Transcript
S ECTION 5 MAINTENANCE AND CALIBRATION PROCEDURES 7. Unscrew the filter mount for the bypass flow channel. 8. Slide the large filter cartridge off the mount and install a new cartridge onto the mount (Figure 5-34). Figure 5-34. Bypass filter mount with clean and dirty filter cartridges. Clean filter cartridge Dirty filter cartridge 9. Install the mount into the unit, then install the cover for the bypass flow. 10. Plug in the sample pump and return to normal operation. THERMO FISHER SCIENTIFIC OPERATING GUIDE, TEOM 1405-D 5-25