Download 1405-D - User Manual
Transcript
S ECTION 5 MAINTENANCE AND CALIBRATION PROCEDURES Calibrating the Ambient Pressure Perform the ambient air temperature calibration, pressure calibration and leak check (Section 3) before executing the flow calibration procedure. To calibrate the ambient pressure: 1. In the TEOM Data screen, select the Service button to display the Service screen, then select the Calibration button to display the Verification & Calibration screen (Figure 5-49). 2. Select the Amibent Calibration button to display the Ambient Calibration screen (Figure 5-54). Figure 5-54. Ambient Calibration screen. 3. Determine the current ambient pressure in atmospheres (absolute pressure, not corrected to sea level). 4. If the measured value is within ± 0.01 atm of the pressure displayed in the Ambient Pressure: field of the button, no further action is necessary. Select the < Back button to return to the Calibration screen. If the value is not within ± 0.01 atm of the pressure displayed in the Ambient Pressure: field of the button, select the Ambient Pressure: button. A keypad will display. Enter the actual pressure as measured by the external device and press the Enter button. The Ambient Pressure Calibration screen will display with the new entered value. Select the < Back button to return to the Calibration screen. THERMO FISHER SCIENTIFIC OPERATING GUIDE, TEOM 1405-D 5-33