SPI – C 170 CHLORINE If the value is correct, select and press enter ( ) on Begin 0.52 mg/l cal. The menu of figure 10.1.2.3 now appears. Cal at 0.52 mg/l Value : 0.51 mg/l Expected: 0.52 mg/l Press(V)to save Figure 10.1.2.3 Supply valve Drain valve Colour valve Remeasure > > > > 6 7 Rinse the cell a few times by alternating the use of the supply valve and the drain valve. Fill the cell with your chosen control fluid. You can either simply pour it gently in the filling tube, or injecting it with a tool. Now select Remeasure and press enter ( ). The cell will now start measuring the control fluid. The value needs to show a result close to 0.52 mg/l. If this is not the case, try to recalibrate again. Otherwise, select Press (V) to save, and press the tick () to save the calibration. 8 The measurement is now calibrated. ! Start calibrating the measurement only after at least one measurement has been performed. Otherwise this could cause a wrong calibration. V1 20130904 19