3 Advanced Troubleshooting
Replacing the Boards in the TSQ 8000 Instrument
5. Locate the PC communication board, which is attached to the controller interface board.
Note Although the PC communication board is attached to the controller interface
board, they are sold separately. See the TSQ 8000 Spare Parts Guide for ordering
information.
.
Figure 325. Locating the Ethernet Cable Connector on the PC Communication Board
Ethernet Cable Connector
6. Disconnect the ethernet cable from the PC communication board.
7. Use a T20 Torxhead screwdriver to loosen the two M4 screws holding the controller
interface board support bracket to the bottom of the instrument.
Thermo Scientific
TSQ 8000 Hardware Manual
299