G Testing
9 Result Page 1/2
10 Result Page 2/2
Picture 10: Printed result
The pad results are displayed on the first page. Positive findings are clearly
marked with red text on the display. To view the remaining test results, touch
the Right
icon on the screen.
The printout is light-sensitive and may turn yellow when exposed to light
during storage. Test results which diverge from negative or normal values are
flagged with an asterisk before the parameter concerned. The printout can be
fully customized, see K.3 Printout for more details. For archiving purposes
the printouts should be kept in a dark place (patient file) or as a photocopy.
Functions on the result screen
•
By pressing Delete
button the result can be dropped.
•
By pressing the Printer
•
By pressing the Transfer
•
To go back to the Measurement, press the Meas. button.
button the result can be printed.
button the result can be transferred.
How to modify the result?
Results can be modified by pressing the Edit button on the second result page,
before the record is printed or sent.
All fields can be modified except date and pad results, even if the particular
field was not available during the acquisition.
Before performing the next measurement
From the test table, remove the used urinalysis strip and dispose of it
according to your standard laboratory procedures. Wipe the table insert, if
necessary.
G.2
Test features and customization
The testing process can be customized to the need of the laboratory. The
measurement feature settings define what activities related to testing process
(start) are automatically performed by the instrument. The measurement
31 / 85