2.5.4.2 Leica Lithography Systems
Electron beam systems from Leica Lithography Systems Ltd. (LLS) are a combination of products
previously manufactured by Cambridge Instruments, the electron beam lithography division of Philips,
and most recently products from the former Jenoptik Microlit Division. Leica sells eight different
models of Gaussian spot vector scan machines (the EBL Nanowriter has been described above).
Systems in the mid-range of resolution include the EBML-300, a LaB6 tool directly evolved from the
Cambridge line, and the EBPG-5, a LaB6 machine evolved from the Philips line. The EBPG-5 is
comparable to the JEOL JBX-5DII in resolution but has accelerating voltage up to 100 kV. The
EBMLand EBPG are both known for their versatile control software. On Leica's high end is the
VectorBeam, with optics evolved from the Philips EBPG line and control electronics and software
evolved from the Cambridge EBML line. The VectorBeam (Fig. 2.19) has a thermal field emission
electron source running at 100 kV and a 6 in. stage motion with up to /1024 = 0.6 nm precision. The
25 MHz pattern generator has the useful feature that it is able to hold a small pattern in a buffer, so
that repeated patterns do not have to be retransmitted to the pattern generator. This can significantly
decrease the transmission overhead time when writing a large array of simple figures.
FIGURE 2.18 Probe beam diameter versus current for (a) a LaB6 cathode with a 120 um objective
aperture, (b) a thermal field-emission (TFE) cathode with a 40 um objective aperture, and (c) a thermal
field-emission cathode with a 100 um objective aperture. Data is from JEOL Gaussian-spot e-beam
systems using 50 kV acceleration and a short working distance objective ("5th lens") (courtesy of JEOL
Ltd.).
Leica e-beam tools are also distinguished from those of JEOL by their use of a single objective lens
(one working distance), and scaleable writing fields with 215=32768 or 216 =65536 pixels across the
field. In the case of the EBML-300, field sizes up to 3.2 mm may be used, although the benefit of using
such a large field is debatable.