Download - Frank`s Hospital Workshop
Transcript
DIAGNOSTICS REPLACEMENT PROCEDURES 6 Discard the old lamp assembly. IMPORTANT Risk of compromising output of the new lamp if the surface is smudged. Fingerprints or other smudges on the lamp can affect output. Do not touch the surface of the lamp. 7 8-32 Using care not to touch the surface of the lamp: a. Insert the new lamp assembly inside the housing. b. Place the bracket (with wings up) on the housing. c. Turn the lamp assembly clockwise until secure. d. Reinstall the two screws removed in step 5. e. Reconnect the lamp to the Power Supply. 8 Plug the instrument’s power cord into the electrical outlet. 9 Verify correct operation: a. Turn the instrument on. b. While the instrument is performing a startup and background check, look to see that the new lamp is lighted. r If it is, go to step 10. r If it is not, then troubleshoot the system to determine the problem. D B A PN 4237615B