Option <2> ‘PR TimeThreshold’ is used as a deployment termination condition. This
defines how long the Profiler travels at the PR Threshold dBar/sec rate before terminating the
deployment. The firmware ends the profile if the speed of movement on the wire falls to the PR
Threshold rate (plus or minus the shallow/deep rates set in the deployment parameters) for the
duration specified here.
Selection [] ? 2
Pressure rate time threshold [sec] (5-300) [180] ?
Figure 3-6: Pressure Rate Time Threshold
Option <3> ‘Sensor warmup’ changes the sensor warmup time from the default which is
120 seconds.
Selection [] ? 3
Sensor warm-up interval [sec] (0 to 300) [120] ?
Figure 3-7: Sensor Warm Up Interval
Option <4> ‘Sensor warmdown’ changes the sensor warmdown time from the default
which is 120 seconds.
Selection [] ? 4
Sensor warm-down interval [sec] (0 to 300) [120] ?
Figure 3-8: Sensor Warm Down Interval