Trouble shooting 15.16. Separator error If it takes too long for the waste pump to empty the liquid separator, the system generates a separator error. Separator error may be caused by: Extensive foam build-up in the liquid separator. Check the separator assembly and connections for possible air leaks. Waste-tube between liquid separator and Replace the tube. waste pump is blocked. Waste-tube between waste pump and waste container blocked. Replace the tube. Waste pump failure. Exchange the waste pump cassette. If the error returns, call for service. Electrical bridge between the waste-level Clean liquid separator, see WI-196 Cleaning liquid electrodes. separator (on page 304) Page 188 StaRRsed Flex User Manual Version 1.02 MRN-151-EN