Download StaRRsed Flex User Manual

Transcript
Appendix for StaRRsed Flex
Appendix - Sysmex R-3500 Protocol
R-3500 sample data record format (202 bytes)
This is a modified data record coming from a R-3500.
Sample data record format (202 bytes)
Parameter
Text distinction code I
# Of
chars
1
Example Comment
"D"
Text distinction code II
1
"1"
Sample distinction code
1
"U"
Day
2
23 Day 23
Month
2
03 Month 3 = march
Year
2
00 Year 00 = 2000
Rack no.
4
Tube position no.
2
Sequence no.
5
ID information
1
Sample ID number
13
Patient number
Instrument ID number
9
ID number from compact
Analysis information
1
0 n.a.
Reserved
18
0 n.a.
RET%
5
12300 Esr value = 123
RET#
5
00200 Hazy code = 2
RBC
5
00200 Error code = 2
IRF
5
01200 Temperature in degr. Celsius = 12 degr.
LFR
5
01200 Sedimentation time in minutes = 12 min.
MFR
5
10100 Dilution rate = 101
HFR
5
12300 30 minute ESR value = 123
Reserved
105
1234 Rack number = 1234
05 Tube position in rack = 5
00000 n.a.
4 Barcode from barcode label
0000… n.a.
n.a. = not applicable
StaRRsed Flex User Manual
Version 1.02 MRN-151-EN
Page 259