If a calibration pack is available on the installation CD received from XenICs, the Camera Setup installer will install this calibration pack in the calibration pack folder. The first time the X-Control software is started, the calibration will be loaded. So by default, your camera is calibrated by XenICs and the calibration pack is installed. The calibration pack is unique for every detector chip and related to all applied camera settings. Therefore the user shall use a calibration pack entitled to a given sensor with specific settings. You may need to execute a calibration procedure whenever the application of the camera usage requires different settings compared to the settings used by XenICs. The calibration is performed with X-Control and generates a related calibration pack. X-Control User Manual 36 June 22, 2006