Download SLM Utility G3 Manual
Transcript
Measurement Settings For a detailed description of the settings defined on this tab, refer to the instrument’s user manual. The settings available on the Measurement sub-tab define some basic sound level parameters such as Frequency and Peak Weighting, Detector type, Integration Method, Ln percentages, as well as OBA (Octave Band Analyzer) settings such as Range, Bandwidth, Weighting, and Spectrum as shown in FIGURE 3-8. Octave Band Data Note that the OBA settings will only be available if the appropriate optional firmware has been installed: • Model 831: 831-OB3 (1/1 and 1/3 octaves) • SoundTrack LxT: LxT-OB1 (1/1 octave only) or LxTOB3 (1/1 and 1/3 octaves) Dosimeter Settings In order to measure noise dose, the Model 831 must have the optional firmware 831-IH loaded. This functionality is standard with the SoundTrack LxT. The settings defining the measurement of noise dose are set from the Dosimeter sub-tab, shown in FIGURE 3-9. Predefined Setups The Dosimeter settings can be loaded from a set of Standards based pre-defined criteria by clicking the Load Std. Criteria button which will display the Standard Dose Criteria dialog as shown in FIGURE 3-10. Simply highlight the desired Standard and click OK to automatically populate the settings. Measurement Control Settings For a detailed description of the measurement control settings, with and without time history, refer to the instrument’s user manual. The settings available on the Measurement Control sub-tab, as shown in FIGURE 3-11, will define how and when the instrument will RUN, STOP and/or STORE data when taking a measurement and if a Measurement History will be kept. Most of the settings on this tab are shown/hidden and/ or enabled/disabled based primarily on the Run Mode setting and the Enable Measurement History setting. Measurement History If an FST Period is selected on the Time History sub-tab, the Enable Measurement History setting will be disabled. SLM Utility-G3 Manual The Enable Measurement History setting determines if a measurement history will be saved and stored with each Measurement. Of the remaining settings on this sub-tab, only "Interval Time Sync." and "Interval Time" are dependent on this setting. Settings Tab 3-21