Working Temperature: Sets working temperature for reading cuvettes and activates the refrigerator for the reagents. Click to select the desired temperature among the available (Figure 13). • Room Temperature • 32°C • 37°C “Enable Reagent Refrigerator”: If not selected, it excludes the reagent refrigeration. Figure 13 “Samples’ Tray”: Customizes samples’ tray (Figure 14). Here it is possible to change the number of positions dedicated to the samples of Routine, STAT, Standards and Controls. Routine and STAT positions are included in the two outermost rings with progressive numeration from 1 to 52. Standard and Controls positions are in the innermost Figure 14 rings with progressive numeration from 1 to 26. It is not possible to exclude one typology to advantage of the other one. The minimum number of assignable positions is "1" for the Routine, STAT and Standards; and "3" for the controls. Click up/down arrows to increase/decrease the number of dedicated positions, or write the desired number into the textbox. The analyzer will automatically upgrade the positions for the second item of the pair every time the first one is changed. “Passwords”: This function is for personalization of the passwords that grant access to program levels 2 and 3 (Fig. 15). This operation is reserved for specialized personnel only. Figure 15 Section I Chapter H Analyzer Technical Functions/Setup BT3000 PLUS Rev.0, Soft Ver. 8 Page 10 of 11