USING VOLTAGE AND CURRENT TRANSDUCERS
REMOVE ANY CONNECTIONS TO THE NORMAL ALO TERMINAL!
!
EXT-LO and ALO are connected inside the PM1000+. To avoid errors
and a risk of electric shock, remove all connections to ALO.
3. Set up the PM1000+ to measure current from the EXT-HI and EXT-LO terminals.
Press ’MENU’
Select  ‘Inputs’ and press 
Select  ‘Shunts’ and press 
Select  ‘External’ and press 
Press ‘MENU’ to return to the measurement display.
4. Scale the measurement on the display.
The default scale is 1V = 80A.
The scale required for any external shunt is given by:
Scale = 1 / (80 x R)
(where R is the resistance value of the external shunt).
In this example where R = 0.001 Ohms
Scale = 1 / (80 x 0.001)
Scale = 1 / 0.08 = 12.5
To enter a scale factor for current:
Press ’MENU’
Select  ‘Inputs’ and press 
Select  ‘Scaling’ and press 
Select  ‘Amps’ and press 
Page 24
© Voltech 2011
PM1000+ Power Analyzer User Manual