Download Chromeleon/MSQ Plus Operator`s Guide
Transcript
Report Definitions ____________________________________Mass Spectra Data Extraction [Integration and QNT-Editor Views] 7. Report Definitions There are limitless options for arranging the data in the printer layout screen. It may be easiest to start with the Dionex template. 1. Open a data file. 2. From the “Workspace” menu, select “Load Report Definition”. 3. Browse the directories until you reach the locked Dionex Templates folder. Open this folder and then select “Reports”. Open “defltLCMS”. 4. Click “No” to the option to save printer layout changes. 5. Go to the View menu and select “Printer Layout”. 6. Scroll through the tabs at the bottom of the screen to discover all of the different views. 7. To customize a view: a. First the report definition must be saved outside of the Dionex Template folder. This is done by right clicking on the page and selecting “Save Report Definition”. SAVE THE FILE OUTSIDE OF THE LOCKED DIONEX TEMPLATE FOLDER. b. The chromatogram can now be customized as desired. c. To change the columns displaying integration values, you must enter Layout Mode. _____________________ Chromeleon / MSQ Plus Operator’s Guide ____________________ 7-1