Microcyber Inc 9.Technical Specification 9.1Basic Parameters Resistance: Input Signal PT100, CU50, CU100, 0~500Ω, 0~2000Ω Thermocouple: B, E, J, N, K, R, S, T Voltage signal: -100mV~100mV Channels 4 Channels RTD connection 2, 3 wire Power Supply of Fieldbus Insulation FF/PA:9~32 VDC/Current : ≤ 14mA Between housing and terminal board: 500 Vrms (707 VDC) Display 6-digit numeric and 5-digit alphabet LCD Display Operation -40 ~ 85℃ (Without display) Temperature -30 ~ 70℃ (With display) Humidity Range 0% ~ 100% RH Start Time ≤ 5s Update Time 0.2s Protection grade IP 65 EMC Designed to comply IEC 61000 (GB/T 18268-2000) 9.2 RTD Specification RTD Accuracy (25℃) Sensor Type Working Range(℃) - 84 - Accuracy