10.2.2 Sensor Maintenance Water and ice must be kept off the sensor in order to have reliable readings. If ice forms on the electrical connectors at the top of the dewar, the readings may not be accurate. 10.3 Bed Maintenance The bed must be lubricated every three months and the belts checked for tension. Care--Watch for the pulley screw on the motor if power is applied and the motor turns. This is not recommended. Do not apply power if the sides are uncovered. 10.4 Water Bag Maintenance The water bag must be checked for leaks periodically and the system kept free of trapped air. Maintenance of the water bag is covered in a separate document. 10.5 Mechanical Pump Maintenance Check the pump for backstreaming and oil periodically. 10.6 Transfer Tube Maintenance The two piece flexible transfer tube has an evacuated region in each section between the inner tube and the outer tube which is at room temperature. If the vacuum deteriorates, then the thermal isolation between the two tubes will also deteriorate, and the outer tube will frost up along its whole length when a transfer is attempted. It is possible that it will be impossible to transfer any liquid helium. The cause of such a deterioration may be either a leak or the mistaken opening of the evacuation valve. If the problem is a leak, it must be fixed. After repair, the transfer tube vacuum space must be evacuated before the transfer tube can be used. As little as 1/2 hour of pumping with a mechanical vacuum pump should be sufficient. 11 Appendices 11.1 Hardware PC and National cards setup Channel list is shown below: Channel assignment NI 6031 (64 Ch) 100 pin Callout ribbon AI SQUID A 3-4 A0,8 1 2 SQUID B 5-6 1,9 3 SQUID C 7-8 2,10 4 H20 9-10 3,11 Pressure 5 Z Position 11-12 4,12 6 Cardiac 13-14 5,13 7 X fluxgate 15-16 6,14 8 Y fluxgate 17-18 7,15 Model 5700 Biosusceptometer page 64 of 76 7/3/99