Installation Bench Acceptance Test Bench Acceptance Test Prior to installing the Model 5030, you should perform the acceptance testing procedures. These tests are conducted to evaluate the out-of-box performance of the instrument, perform any necessary calibrations prior to final site installation, and familiarize the user with the menu structure. Use the following procedure to perform the bench acceptance test. Equipment Required: Model 5030 SHARP Monitor, Model 5030 Power cord Mini-ambient/heater temperature cable assembly (DO NOT USE THE HEATED SAMPLE TUBE ASSEMBLY AT THIS POINT) Vacuum Pump Assembly Flow Adapter Assembly NIST-traceable Thermometer NIST-traceable Hygrometer NIST-traceable Barometer NIST-traceable Volumetric Flow Transfer Standard 1. After unpacking the instrument place the Model 5030 and accessories onto a table or bench located within a stable indoor environment. Prior to continuing with acceptance testing, the instrumentation and accessories should have sufficient time to equilibrate to room temperature due to temperature variations during shipping and/or storage. 2. Following a sufficient equilibration period, connect the power cord to the rear of the instrument and to a properly grounded power supply. If the Model 5030 turns ON at this point, set the power switch to the “OFF” position and then proceed. 3. Connect the braided vacuum hose to the top brass port on the pump and the other end to the vacuum pump hose connection on the rear of the monitor. Thermo Fisher Scientific Model 5030 SHARP Monitor Instruction Manual 2-7