Acuity Staircase Test
expected to be below the 1% figure, and another 1% above the 99% figure.
Of course this means that 98% of the population is expected to be between
these boundaries. In this starting case, the display is showing a 98%
probability that the starting contrast will be somewhere between the ratios
of 6/3.12 and 6/116.23. As this is almost the entire probability range at this
point, it is a fairly safe bet given that no test has yet been made.
With the operator pressing the yes/no keys in response to the patient’s
affirmative or negative replies, the test will proceed until the 1% to 99%
boundaries contain less than 5% of the initial range. At this point the test
ends with the handheld emitting three short beeps and displaying a final
screen similar to this:
The Letter is the final seen letter. The Acuity figure is the final Snellen
fraction – the test result. The 1% and 99% figures define the boundaries of
a range of Snellen fractions within which the final figure could appear with
a 98% probability. The test result can be anywhere between these
boundaries. The smaller this range the higher the confidence that can be
placed in the final figure.
56