The comparative method used in regressions was BD FACSCalibur with BD
Multiset™ software and BD Multitest™ IMK Kit with BD Trucount tubes.
•
The least-squares method was used to fit the linear regression for all
subsets.
•
Accuracy studies were conducted over a period of 12 days at two sites.
•
Number of replicate determinations used to calculate each mean for X: two
replicates for CD3, one for all other subsets.
•
Number of replicate determinations used to calculate each mean for Y:
minimum of two and maximum of four replicates for CD3 and minimum
of one and maximum of two for all other subsets.
BD FACSCanto Flow Cytometer Reference Manual