Download LTQ Hardware
Transcript
3 System Shutdown, Startup, and Reset Turning Off Selected Mass Spectrometer Components Table 5. On/off status of mass spectrometer components, voltages, and gas flows (Sheet 1 of 2) Mass spectrometer component Vent valve Standby mode Off mode Electronics service switch, Service Mode position Main Power switch, Off (O) position Closed Closed Closed Open (5 sec) APCI corona discharge needle APCI vaporizer Conversion dynode Electron multiplier ESI needle Gas, auxiliary Gas, sheath Off Gas, sweep Ion optics multipoles rf voltages Off Mass analyzer, rf/waveform voltages Power supply, 8 kV Power supply, electron multipliers/conversion dynodes Off API ion transfer tube dc offset Off API tube lens Ion optics lens Ion optics multipoles dc offset voltages Mass analyzer, dc offset voltage Power supply, 300 Vdc API ion transfer tube heater Fan, rf coil On Gauge, Convectron Gauge, ion Power supply, PS1 On Fan, turbomolecular pump Fans, electronics tower Forepump On Power supply, PS2 42 LTQ Series Hardware Manual Thermo Scientific