of the slopes of the larger peaks. To access this function again, go to the “Preparation” drop down menu, and this time select “Line shape analysis” located right under the “Define noise level” function. Once selected, this function will open a window asking you to confirm the noise level number. The number it displays should be the number that you had selected as noise level before. (Screenshot 10) Confirm by pressing “OK” and Amix will clean up all the noise below your selected noise level. Your spectra is now almost ready to be saved. There are a number of functions for editing areas of the spectra which are Screenshot 10: The Noise threshold number recived from the “Define noise level interactivly” function unwanted, such as peaks from water or other solvents. You can remove these peaks by using the “Delete area” function in the “Preparation” menu. This function will allow you to highlight an area on the spectra using your mouse. This area will be deleted from the spectra. There are also functions such as “Remove Positive” or “Remove Negative” which will delete all positive or all negative peaks from your spectra. The “Keep area” function will ask you to highlight an area which should not to be deleted, the rest of the spectra will be deleted. After removing all areas which are not relevant the spectra can be saved to Sbase. To do this select the “Save to spectral base” function in the “Preparation” drop down menu. A window will appear which will allow you to define the compound and the experiment type this spectra should be saved to. (Screenshot 11) 75