Download TidalGuard Sp
Transcript
Preparation for Use Battery Use and Options Charging NiMH Rechargeable Battery with External Charger In a non-patient area, connect the adapter to an AC source, then plug the adapter jack into the charger. Remove the battery from the TidalGuard™ Sp and insert it into the external charger. The battery will be fully charged in approximately 4.5 hours. The external charger is for use with the DR30 NiMH rechargeable battery pack (Cat. No. 400043) only. Refer to the instructions supplied with the charger for additional information. DR30 NiMH Rechargeable Battery (Cat. No. 400043) Connect adapter Battery is keyed to fit into slot in only one direction Connect line cord WARNING • The external battery charger should NOT be used to recharge the battery near or in close proximity to patients and/or other medical equipment in operation. It is intended for use in service areas only (i.e. nurses station, biomed lab, etc.). NOTE • • • • Rev. B With a new battery, or a battery that has not been used for 30 days or more, charge the battery for 24 hours prior to use. When powered by the external power supply or the BaseStation, the TidalGuard™ Sp will not overcharge a rechargeable battery. The monitor may not operate on battery power if the battery is not sufficiently charged. Dispose of batteries in accordance with local laws. TidalGuard Sp User’s Manual 13