BlazerX6 User’s Manual
Integral Solutions Int'l
September 3, 2013
Valid Head Resistance
Range of acceptable resistance, for detecting point of first
contact.
Maximum Distance
Total slider will move right during Find First Contact
procedure before stopping if resistance measured after each
step was not acceptable.
Step Size
Distance slider will move right during Find First Contact
routine to measure resistance. The slider will keep moving
right until maximum distance is reached.
Find Point Of First Contact
Use this function to detect the point of first contact of the
probe card with the slider. The function will start from current
W position and move right one Steps Size at a time. After
each step, the routine will measure resistance and stop if
resistance is in the acceptable range. Otherwise the function
will continue, until slider has moved maximum distance from
the position where this function was initiated. The procedure
measures resistance on both channels simultaneously and
stops if resistance is valid on either one. If it finds the first
contact, then it reprobes 5 times and measures resistance.
Then software will prompts 5 resistance readings, so you can
decide how good is the contact. You can then find more
precise position where both sliders are making contact and set
probes position parameter. Power On checkbox should be
enabled and I+I- pin pair should be used.
Movement will start
without confirmation in
service mode. Keep hands
clear.
Probes Location
Position where the probes would touch the wall of the Wstage pocket. User should never type in such position, or try to
detect it. Always put sliders in pockets, and detect point of
contact, then software can calculate this position based on the
slider size.
Overdrive
Amount of overdrive software should use when probing. [60]
0 Point Offset
Distance the software will move back from the point of first
contact when user clicks executes Unprobe function. The
same distance will be used during Find First Contact routine
to remeasure resistance 5 times.
Calculate Probes Location
Takes current W-Stage position and subtracts slider length to
find Probes Location. You must input slider length prior to
running this function. Slider should be at the point of first
contact, ie the first position where both sliders are making
contact with the pins on the probe card. Before calculating
probes location, always check that both I+I- resistance and
MR+MR- in Up and Down configuration on both channels –
all read valid resistance. If not you may have to realign probe
card or move sliders closer to probes.
Schedule Maintenance
TBD