Download BlazerX5/X6 Slider Configuration Supplement Integral Solutions Int`l
Transcript
BlazerX6 User’s Manual Integral Solutions Int'l September 3, 2013 Valid Head Resistance Range of acceptable resistance, for detecting point of first contact. Maximum Distance Total slider will move right during Find First Contact procedure before stopping if resistance measured after each step was not acceptable. Step Size Distance slider will move right during Find First Contact routine to measure resistance. The slider will keep moving right until maximum distance is reached. Find Point Of First Contact Use this function to detect the point of first contact of the probe card with the slider. The function will start from current W position and move right one Steps Size at a time. After each step, the routine will measure resistance and stop if resistance is in the acceptable range. Otherwise the function will continue, until slider has moved maximum distance from the position where this function was initiated. The procedure measures resistance on both channels simultaneously and stops if resistance is valid on either one. If it finds the first contact, then it reprobes 5 times and measures resistance. Then software will prompts 5 resistance readings, so you can decide how good is the contact. You can then find more precise position where both sliders are making contact and set probes position parameter. Power On checkbox should be enabled and I+I- pin pair should be used. Movement will start without confirmation in service mode. Keep hands clear. Probes Location Position where the probes would touch the wall of the Wstage pocket. User should never type in such position, or try to detect it. Always put sliders in pockets, and detect point of contact, then software can calculate this position based on the slider size. Overdrive Amount of overdrive software should use when probing. [60] 0 Point Offset Distance the software will move back from the point of first contact when user clicks executes Unprobe function. The same distance will be used during Find First Contact routine to remeasure resistance 5 times. Calculate Probes Location Takes current W-Stage position and subtracts slider length to find Probes Location. You must input slider length prior to running this function. Slider should be at the point of first contact, ie the first position where both sliders are making contact with the pins on the probe card. Before calculating probes location, always check that both I+I- resistance and MR+MR- in Up and Down configuration on both channels – all read valid resistance. If not you may have to realign probe card or move sliders closer to probes. Schedule Maintenance TBD Page 26