The Autofocus calibration moves the stage approximately 5 microns out of ideal focus and collects focus error scores from 100 SpeedBeads. The stage then moves one micron toward focus and collects focus error scores again. The process is repeated until the stage reaches a position 5 microns beyond best focus. The data are then fit to a 3rd order polynomial and the slope of the curve at the inflection point is determined. This value determines the signal error corresponding to a given focus error through the typical operating region of the autofocus system. It is used to determine each move the focus motor makes to correct focus errors. The result and the limits for the calibration are shown below the list when the calibration is selected. If the Multimag option is installed the instrument automatically multiplies the S curve slope by the appropriate factors to compensate for the changes in slope that result from the additional 60X and 20X magnification. - 69 -