INSTALLATION AND SETUP
MODEL 700
3.4
3-33
SETTING THE COM ID
The Model 700 Com ID is determined by dual
inline package (DIP) switch settings.
Follow the steps in this
section only if you wish to
do the following:
• Change the Com ID
setting.
• Visually inspect and verify
the DIP switch settings.
Figure 3-15 Dip Switch
In most cases, the Com ID configuration made
at the factory will not need to be changed. The
factory DIP switch settings produce a Com ID
of 1.
3.4.1
Note that the MON2000
software Com I.D.s will
override the hardware
settings. To use the
hardware Com I.D.
configuration, enter a zero
‘0’ in the MON2000 Com
I.D, settings. For details,
refer to MON2000 Software
for Gas Chromatographs
User Manual (P/N 3-9000522).
Inspect or Change the Com ID
To visually inspect and verify the DIP switch
settings or to change the Com ID settings on
the multifunction board,
1. Remove the dome from the upper enclosure.
MAY CAUSE INJURY TO PERSONNEL OR
EQUIPMENTDAMAGE
Use caution when accessing an enclosure. A
voltage of 115 to 230 VAC, along with various
DC voltages, are present.
Failure to follow this precaution may cause
injury to personnel or damage equipment.
2. Loosen the thumb screws on the Oven
System mounting plate.
JULY 2007
Setting the Com ID