Transcript
SPECIFICATIONS 3. Specifications 3.1 Environmental Requirements It is essential that the climate of the laboratory is controlled to ensure that the CCD detector is not damaged. Typically air-conditioning should be installed to maintain the temperature and humidity within the ranges listed below. The Relative Humidity is particularly important as the CCD and its cooling pipes can reach ~17C, condensation should not be allowed to collect on the CCD at any time. Xcalibur/Gemini with KMW3000 (i.e. external water cooled) dissipates ~6000 BTU/hr heat, not including any optional cryogenic devices. Xcalibur/Gemini with TF3500 (i.e. air cooled) dissipates ~16000 BTU/hr heat, not including any optional cryogenic devices. Air temperature in the room during operation. 18 – 25 C Stability of ambient temperature during operation 1 C Storage temperature >10C <40C Relative humidity 20 - 80 % non – condensing Location Clean, dust free environment >2m from air conditioning or heating units Floor strength Able to bear system weight of 500 kg on a footprint 80x80cm Xcalibur Version 2.4 USER MANUAL Page 16