Download User Manual - Agilent Technologies
Transcript
MAINTENANCE INSTRUCTIONS 8.3 Checking the Door Safety Interlocks Task Time: 2 minutes When: Once a week Tools: None Procedure WARNING Ensure the interlocks are not defeated. If interlock defeated there is the risk of exposure to X-ray radiation. With the diffractometer in normal operation and at normal X-ray generator settings, with the shutter open, open the front door of enclosure cabinet. The X-ray shutter should immediately close. Close the enclosure door. From the software open the shutter again. Open the left side panel of the cabinet. The X-ray shutter should immediately close. Close the side panel. From the software open the shutter again. Open the right side panel of the cabinet. The X-ray shutter should immediately close. Close the side panel. From the software open the shutter again. Open the back panel of the cabinet. X-ray shutter should immediately close. Close the back panel. Record the date, persons testing and sign off the outcome. 8.4 Checking the Emergency stop Task Time: 10 minutes When: Once a month Tools: None Procedure 1. With the diffractometer running at minimum settings (for example 10 kV, 0.05 mA), press red Emergency Stop button. 2. The x-ray generator should switch off immediately, release the emergency stop button, restart the x-ray generator. Return to required settings. 3. The Crysalis software will indicate that the emergency stop was activated: switch off the interface, restart the interface and then type command gon reinit in order to restart the system. 4. Record details of the test in the local radiation log, stipulating the date, person testing, outcome and signature. . Xcalibur Version 2.4 USER MANUAL Page 51