TESTS AND TEST CONDITIONS 7.1.4 R2 – DC Resistance Match Measurement Range 1:1000 to 1000:1 Test Voltage N/a AT Series Tester Test Frequency N/a Basic Accuracy 0.2% R1 R2 Where Used The DC resistance match test – as opposed to an ordinary DC resistance measurement (R) - is used on audio and telecommunications transformers, where it is important that the resistance of different pairs of windings is controlled and matched to a specified ratio. The absolute value of the resistances may be of less importance to the performance of the transformer than the match between two resistances. Test Conditions To measure DC resistance match, the tester makes two DC resistance measurements (see the R test) and compares the two results. Limits for the match of the two measured resistances may be set in terms of the ratio between them (e.g. 1:1 ± 5%). By adding further DC resistance match tests to the test program, any number of DC resistances can be tested for match. PAGE 7-16 VOLTECH ATI USER MANUAL