Download ATi USER MANUAL
Transcript
TESTS AND TEST CONDITIONS 7.1.5 LS, LP – Inductance (Series and Parallel Equivalent Circuit) Range Test Voltage Test Frequency Basic Accuracy 1nH to 1MH 1mV to 5V 20Hz to 3MHz 0.05% Ls AT Series Tester Rp Lp Rs Inductance is the electrical parameter that is commonly used to check that a high-frequency transformer has been constructed according to its design and that it will operate properly in use. The inductance of a winding is determined by the number of turns and the construction and dimensions of the core materials; it may also vary with test conditions such as voltage and frequency. The series and parallel inductance tests are used when testing signal, pulse, switched mode and similar transformers with ferrite type cores, where the inductance remains the same from small signals up to the operating conditions. Faults in numbers of turns, core material and core gapping may be determined using this test. For line frequency iron cored transformers where the inductance varies rapidly with voltage a magnetizing current test (MAGI) is recommended. (The MAGI test is only available on the Voltech AT3600). VOLTECH ATI USER MANUAL PAGE 7-19