Chart 3-1B: Connection of Two Cables
If many series cables in different locations appear in on system,
a one time series calibration is suitable.
When testing a power amplifier, connect an attenuator with the
output port in series to avoid damage. If it is a little signal
amplifier, connecting an attenuator with the input port in series
is imperative to avoid strong signal. Before that kind of
measurement, it is suggested that all the attenuators should be
connected with cables in series to operate calibration.
Follow the hints of ReCAL in function menu to start user calibration.
If frequency is changed after calibration, the system calibration
parameters are automatically used.
3.5 Calibration before S11 Measurement
Although there are many display modes in S11 function, only one
calibration is necessary, because the same data is being shared in
all modes.
It is necessary to operate user calibration in following cases:
1. In acceptance measurements for important antennas, for example,
antennas of radio and TV emission stations.
2. Failing to be connected to KC901S, the port under measurement