Transcript
Coulometrics User’s Manual 10 Weighing Carbonate Samples Carbonate samples are weighed using the Cahn balance. The balance is connected to a Coulometrics balance program that records the sample mass, calculates a standard error for each sample, and plots each mass graphically. Start the balance program by double clicking on its icon. This opens the “Balance” window (see Figure 14). Figure 14. The “Balance” window. Calibrating the Balance Before samples are weighed, the Cahn balance must be calibrated. Calibrate the balance using the Coulometrics balance progam. Two calibration are done during port call: 0-10 mg for Coulometer/CNS; 1-100 mg for RockEval. The balance should be calibrated with the hardware first and then with the software. Calibrating the Hardware 1. Set the Cahn balance to measure in the 250 mg to 1 microgram range. Ocean Drilling Program