Download HPLC
Transcript
12 Injection Port 12.5 Split/Splitless Injection System 12.5.1.2 Parameter list TEMP Range: 0.0−450.0°C, Default: 25.0°C Set the temperature of the injection port. Keep the injection port temperature as low as is practical to increase the service life of the temperature sensor. In order to instantly vaporize samples inside the injection port, set the injection port temperature approximately 30°C above the column oven temperature (final temp). Due to the minute quantity of sample injected, the sample vaporizes at temperatures lower than the boiling point of sample elements. To set the injection port maximum temperature limit, refer to “16.6.4 Setting the maximum temperature limits” in “16.6 GC Configuration”. 12.5.1.3 PF menu PF menu 126 Description Reference section Print Prints the injection port temperature on Chromatopac. ―― Next Inj Toggles among the setup screeens if two or more injection ports are installed in this system. The [INJ] key can also be used to switch screens. ―― GC-2010