Download HPLC
Transcript
19 Troubleshooting 19.2 Error Messages AOC communication errors Code Message CS 4009 AOC command is invalid No 4010 AOC data is out of range No 4011 AOC time out No 4012 AOC2 TRS error No 4013 AOC link error No Note One of these messages appear during a link failure or communication failure.(When a communication error occurs, the link is automatically disconnected.) When one of these messages appear, check the connection status, and reset the link. 19.2.5 Detector errors Over current Code Message CS Note 4101-4104 DET#n TCD cell error Yes n=1-4 4105-4108 DET#n FTD current error Yes n=1-4 Abnormal detector current. If the resistance of the filament becomes abnormally high when the TCD or the FTD is in use, the protection circuit is actuated to prevent damage to the filament, and an alarm sounds. If either error has occurred, turn off the system. Possible reasons for the protection circuit to be activated are described below. The current value set exceeds the maximum operating current. The makeup gas is not flowing. A gas leak has occurred. A significant amount of air is present in the flow line (for TCD). Remove the cause of error, then turn on the system. If the system does not recover after several resets, or the reason for the error cannot be located, contact your Shimadzu representative. Detecher flame error Code Message 4109-4112 DET#n flame is out CS No Note n=1-4 The detector flame (FID) has been extinguished. Check the gas supply, and ingnite the detector again. If the flame error occurs repeatedly, the hardware has failed. The system cannot be used. Turn off the system, and contact your Shimadzu representative. GC-2010 315