Download HPLC
Transcript
3 AFC, APC 3.6 AFC and APC Offset Calibration Fig. 3.6.2 Perform the offset calibration of the APC using the following procedure. (1) Press[SYSTEM]key, toggle key, PF1 key, and change “Start GC” to “Manual Start” (refer to Fig.3.6.1 or 3.6.2). If “Manual Start” was already set, proceed to the next step. (2) Turn off the power. Wait until the column oven has cooled down sufficiently. (3) Turn on the power. Wait for approximately 1 minute. (4) Press the [DET] key, the [PF2] key and the toggle key, in that order. The screen shown in Fig. 3.6.3 opens. (5) Make sure that flow monitors of each gases do not change. When they change, wait for a while. Press the PF2 key. The message “Zero calibration Start” appears at the bottom of the monitor. Approximately 10 seconds later, the message “Zero calibration completeds” appears. Calibration of the APC is complete. (6) Return the “Start GC” setting to its original status. GC-2010 47