Download Technical Publication
Transcript
GE MEDICAL SYSTEMS DIRECTION 2300164-100, REVISION 7 8-8-5 NOTE: VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL Panasonic VCR Replacement Procedure This section describes the installation and removal procedure of the Panasonic AG-MD-835E (PAL) and AG-MD-835P (NTSC) video cassette recorders. The installation and removal procedures are the same for both models. NOTE: IMPORTANT - Before installation of either Panasonic VCR model, it is necessary to order the Panasonic VCR RS-232 Adaptor (Part No. 066E8200). Figure 8-147 Panasonic Video Cassette Recorder (VCR) 8-8-5-1 Tools No special tools needed. 8-8-5-2 Manpower One person - 30 mins. 8-8-5-3 Preparations 1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation. 2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal Procedure on page 8 - 8. 3.) Unpack the VCR. 4.) Check that the VCR AC voltage corresponds to that of your local voltage. 5.) Verify that the switches on the VCR front switch panel are as shown in Figure 8-148 on page 8-152. If necessary, adjust accordingly. Chapter 8 - Replacement Procedures 8-151
Related documents
Artisan Technology Group is your source for quality new and
GE Vivid 3 Pro / Vivid 3 Manual 3
chapter 4
FB091565 Revision 1 GE Medical Systems Vivid 5 Service Manual
Technical Publication Direction 2300000 Revision 0 GE Medical
LOGIQ Book XP/XP PRO/XP Vet Service Manual
Endoscopic Units
CV-180 EVIS EXERA II Video System Center
Technical Publications Vivid 3 User`s Guide
GE Vivid i user manual
Omnesys Technologies Private Limited. NEST An ECO system for
chapter 2