GE MEDICAL SYSTEMS DIRECTION 2300164-100, REVISION 7 7-4-4-3 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL Manual Lights Test 1) Access the HW Diagnostics menu, as described in the Accessing the Back End Diagnostic Options section, on page 7-68. 2) Trackball to the External Keyboard button and press Select. The Extended Keyboard Test window is displayed, as shown in Figure 7-48 on page 7-73. 3) In the Test Manager tab, trackball to the Lights (Manual) radio button and press Select. 4) In the KB Panel tab, deselect the Select checkbox. Verify that after the Select LED is off, the alphanumeric keyboards LEDs are dimmed. 5) When you are finished, trackball to the Exit button and press Select. 7-4-4-4 Commands History Select the Commands History tab in the Extended Keyboard Test window to view a report of the activity of the extended control panel during the test, as shown below: Figure 7-50 Commands History Tab - Extended Keyboard Test Window 7-4-4-5 General Configuration Select the General Configuration tab in the Extended Keyboard Test window to set the type of log files that is generated, for example an error log or no log at all. Chapter 7 - Diagnostics/Troubleshooting 7-75