Download Technical Publication

Transcript
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7
8-8-5
NOTE:
VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Panasonic VCR Replacement Procedure
This section describes the installation and removal procedure of the Panasonic AG-MD-835E (PAL) and
AG-MD-835P (NTSC) video cassette recorders.
The installation and removal procedures are the same for both models.
NOTE:
IMPORTANT - Before installation of either Panasonic VCR model, it is necessary to order the
Panasonic VCR RS-232 Adaptor (Part No. 066E8200).
Figure 8-147 Panasonic Video Cassette Recorder (VCR)
8-8-5-1
Tools
No special tools needed.
8-8-5-2
Manpower
One person - 30 mins.
8-8-5-3
Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Verify that the switches on the VCR front switch panel are as shown in Figure 8-148 on page 8-152.
If necessary, adjust accordingly.
Chapter 8 - Replacement Procedures
8-151