9.3 Accurate Time Measurements Two deep acquisition memories, each storing 32000 points, provide the unprecedented time resolution of the 9400A. In order to match the time base settings, a maximum of 25000 digitized points are displayed on the screen with a resolution of 500 display points. A compacting algorithm showing all minimum and maximum values ensures that no information is lost in the display of a trace. Time cursors can be positioned accurately on any one of the 500 display points of a compacted trace. The corresponding measurement accuracy is 1/500 or 0.2%. digitized waveform max :" :¯ ,: :¯ ¯ ¯ , : : : ¯ ¯ digitized points ...... " : <)............ O.......... """:.. ........... O.......... : f........ .. : . .; ¯ ~ displayed ¯ $ points compacted display with min max algorithm To improve measurement accuracy, two expansion functions, EXPAND A and EXPAND B are provided to display every digitized point trace. When the expansion factor is such that 500 measured points are to be displayed, every display point corresponds to a digitized point. When the time base requires that less then 500 digitized points fill the screen, the 9400A interpolates using straight line segments between the actual points¯ In the example below, the compacted trace of a 25000 point waveform is expanded by a factor of I00 with EXPAND A and B to provide maximum time measurement accuracy. Under these conditions, each expanded trace displays 250 digitized points. Every other display point is interpolated and the time measurement accuracy is 1/50000 = 0.002%. Getting the Most Out of Your 9400A 9-3