Download 2000 User Guide - Xitron Technologies
Transcript
Calibration and Maintenance 69 This output level may be measured directly by using: A high precision DMM, or; The preferred method of using a nullmeter to monitor the deviation between this instrument’s output level and that of a precision calibrated 10V source. 8. Calibrate to an output level of 0.00000A. Adjust the output level of the 2000 instrument as described in step 2, until the voltage measured is: 0.0 +/–2.5V, i.e., a current level of 0 +/– 25pA. This output level may be measured directly by using: A calibration standard 100K resistor and a high precision DMM, or; A nullmeter to monitor the voltage drop across the standard resistor 9. Calibrate to an output level of 1.0000A. Adjust the output level of the 2000 instrument as described in step 2 until the voltage measured is: 100mV +/– 7.5V, i.e., a current of 1A +/– 75pA. Resistor 2000 Voltmeter This output level may be measured directly by using: A calibration standard 100K resistor and a high precision DMM, or; A nullmeter to monitor the voltage drop across the standard resistor. 10. Calibrate to an output level of 10.000A. Adjust the output level of the 2000 instrument as described in step 2 until the voltage measured is: 100mV +/– 5V, i.e., a current of 10A +/– 500pA.