Download model 8015 - CareFusion

Transcript
Authorized User Mode
Authorized User Mode is a feature that:
•
Combines PC Unit tamper resist feature with Auto-ID
application.
•
Is designed to ensure that only clinicians with a bar code
on their ID badge can program Alaris® System.
•
Is available only if it is enabled in selected profile and there
is an Auto-ID Module attached.
When this feature is enabled, the PC Unit automatically
enables the tamper resist mode upon power on and 5 minutes
after programming is completed. If the system is configured
to do so, the Authorized User Mode can be disabled without
scanning a clinician's ID; press and hold the Tamper Resist
Switch (on back of PC Unit) for 3 - 4 seconds.
To unlock the keypad, the user must scan their ID badge or
use the OPTIONS menu to manually input their ID number.
When a questionable bar code is scanned at the main screen
and the keypad is unlocked, a prompt to confirm the type of bar
code scanned appears. This occurs whether the Authorized
User Mode is enabled or disabled.
To use Alaris® System with Authorized User Mode enabled:
1.
Power on system and associate patient ID (see "Patient
Identification" procedure).
Upon successful entry of patient ID, PC Unit automatically
enables tamper resist feature.
2.
To disable tamper resist, press SCAN key and scan
clinician ID badge.
In a very low battery condition, with less than 5 minutes
of battery time remaining, scanner is disabled. In this
situation, disable tamper resist by pressing Tamper Resist
Switch on back of PC Unit for approximately 2 seconds.
3.
Program infusion.
When no keys have been pressed on PC Unit for a
5-minute period, tamper resist mode is automatically
enabled.
Alaris® System DFU – with v9 Model 8015
Auto-ID Module Section
Programming 6-5