Download model 8015 - CareFusion
Transcript
Alarms and Messages (Continued) Audio Characteristics Type Sound EtCO2 Alarm (HIGH PRIORITY) EtCO2 Alarm (LOW PRIORITY) Note A sequence of five beeps. Variable volume; can be silenced for 2 minutes. One long beep approximately every Variable volume; can be silenced for 2 minutes. 4 seconds. EtCO2 Error (Hardware Detected) A single alarm tone volume. Fixed maximum decibel volume; cannot be silenced. EtCO2 Error (Software Detected) Pairs of long beeps. Fixed maximum decibel volume; can be silenced for 2 minutes. Alarms High Priority Alarm Meaning Response CHANNEL ERROR Hardware failure detected by software. To silence alarm and continue operation of unaffected instrument, press CONFIRM soft key. Replace module, as needed. DISPOSABLE DISCONNECTED Microstream® Disposable removed from instrument during monitoring mode. Attach Microstream® Disposable to instrument. HIGH ETCO2 EtCO2 value is above specified Assess patient condition. Confirm correct alarm limit values are selected. HIGH FICO2 FiCO2 value is above specified limit Assess patient condition. Confirm correct alarm limit values are selected. HIGH RR Respiratory rate is above specified limit. Assess patient condition. Confirm correct alarm limit values are selected. LOW ETCO2 EtCO2 value is below specified limit. Assess patient condition. Confirm limit. correct alarm limit values are selected. 5-28 Troubleshooting and Maintenance Alaris® System DFU – with v9 Model 8015 EtCO2 Module Section