Download Model 2701Ethernet-Based DMM / Data Acquisition System Service
Transcript
B-4 Calibration Reference Model 2701 Service Manual Miscellaneous calibration commands Miscellaneous commands perform miscellaneous calibration functions such as programming the calibration code and date. These commands are discussed in detail in the following paragraphs. :CODE (:CALibration:PROTected:CODE) Purpose To program the calibration code or password so that you can perform the calibration procedures. Format :cal:prot:code '<char_string>' Parameter Up to a 8-character string including letters and numbers. Description The :CODE command enables the Model 2701 calibration procedures when performing these procedures over the bus. In general, this command must be sent to the unit before sending any other comprehensive or manufacturing calibration command. The default calibration code is KI002701. NOTES The :CODE command should be sent only once before performing either the comprehensive or factory calibration. Do not send :CODE before each calibration step. To change the code, first send the present code, then send the new code. The code parameter must be enclosed in single quotes. Example :CAL:PROT:CODE 'KI002701' Send default code of KI002701. :COUNt? (:CALibration:PROTected:COUNt?) Purpose To determine how many times the Model 2701 has been calibrated. Format :cal:prot:coun? Response <n> Calibration count. Description The :COUNt? command allows you to determine how many times the Model 2701 has been calibrated. NOTE Use the :COUNt? command to help you monitor for unauthorized calibration procedures. Example :CAL:PROT:COUN? Request calibration count.