Download Ziehm 8000
Transcript
6 Switching the System On and Off 6.1 Preparing the system Before switching on the system, e.g. after a transport, you must connect the two system components to each other and also connect the entire system to the power supply. WARNING Never connect the monitor cart and C-arm stand when the monitor cart is already connected to the power supply or switched on. Damage to the electronics of the system cannot be excluded if this warning is ignored! To prepare a Ziehm 8000, do the following: Ziehm 8000 • Unwind the coupling cable from the upper cable support on the back of the monitor cart (→ Fig. 2-2, p. 2-7). • Plug the coupling cable connector into the socket located at the side of the C-arm stand and lock the connector by pulling the locking lever upward (→ Fig. 2-1, p. 2-6). • Make sure that a suitable supply voltage is available and that the socket-outlet is properly grounded and fused. • Check the power plug on the monitor cart power cable (→ Fig. 2-2, p. 2-7) and the socket-outlet for compatibility. • Connect the system to the power supply. • Make sure that the inclination of the system does not exceed 5° from the level in operating position. • Put on suitable protective clothing. • DICOM: Connect the DICOM port(s) on the back of the monitor cart to a DICOM network socket using the appropriate cable(s). Ziehm 8000 P_26376/CD_28210 - EN - 01/07/2010 6-1