VMI AB SWEDEN 10.2.5.2. Using the Saved Speed For a new un-loaded spectrum, without speed measurement in time of the data acquisition, the SpeedTool Menu appears as below: Measured item is disabled, because the speed was not measured in time of data acquisition. Saved shaft speed item is also disabled, because no any shaft speed was saved. To set the proper shaft speed for all machines directions, proceed as follow: - Select any spectrum of the same machine. - Manually move the shaft speed cursor near the frequency where you consider that the speed is correct. - From the Main Menu toolbar, press button. In the enlarged spectrum graph, locate the Best shaft position. This is the maximum of the peak, as bellow: - Now press again button to display whole spectrum. From the Speed tools item, select Save current speed and Save to all points item. The proper shaft speed will be back-calculated, using the speed settings in the database, for all the measurements belonging to the same machine and having the same transfer time 67