5730A Operators Manual Calibrator 732B DC STANDARD AC PWR IN CAL CHARGE LOW BAT 10V 1.018V CHASSIS 10V COM 1.018V COM GUARD SERIAL NUMBER hhp027.eps Figure 7-4. 732B External Calibration Connections 5. Enter the true value of the 732B 10V output. The true value is the value printed on the calibration sticker or the standard. 6. If the entered value is not between 9 V and 11 V, an error message is shown. The process can be started over from this point with a calibrated 732B. Push to start the calibration procedure. As the 5730A Calibrator does self-calibration, it will indicate what is happening on the display. When the 6.5 V and 13 V references have been characterized, the display prompts the user to accept or reject the changes that are about to be made to the calibration constants. 7. To reject the changes, touch Cancel. Otherwise, touch Continue to accept and save the changes, and to continue with calibration. 8. Reverse the HI and LO connections at the 732B terminals, and push Continue to continue the calibration process. 9. After this part of the calibration procedure is complete, the Calibrator shows the reference shifts and prompts to continue. Touch Continue. See the manual for the correct connections, as prompted by the Calibrator, and enter the first calibration value. 10. Connect the Calibrator to the 10 kΩ standard as shown in Figure 7-5 and enter the true value of the standard. If the standard is not between 9 kΩ and 11 kΩ, an error message is shown. Start over from this point with a different standard. Push again to continue. 11. When the internal 10 kΩ reference has been characterized, accept or reject the changes that are about to be made to the calibration constant. To reject the changes, touch Cancel. Otherwise, touch Continue to accept and save the changes. This lets the calibration process continue. 7-12