Download The IRIS User Guide
Transcript
APPENDIX V – Reference Dose Rates How to treat radioactive samples (ISIS duty officer x6789). >10μSv/hour Store sample in the lead castle for it to decay. >0.1μSv/hour Store sample in IRIS active sample cupboard with sample record sheet. The sample may NOT be removed from its container. For removal from ISIS contact the duty officer. <0.1μSv/hour The sample is not radioactive. For removal from ISIS contact the duty officer As a guide in the planning of your experiment, given below are typical dose rates for an empty annular can: IRIS - Irradiation time 19.5 hrs - Total beam current 2400uA - Flux on sample 1x107 n/cm2.s Time from removal Dose rate on contact from IRIS sample pit (uSv/hr) (hrs) Cap On Cap Off 0 150 250 1 70 150 3 40 90 9 5 20 Dose rate at 10cm from sample (uSv/hr) Cap On Cap Off 8 20 3 9 2 4.5 0.5 1 For comparison: OSIRIS - Irradiation time 14.5 hrs - Total beam current 1800uA - Flux on sample 2x107 n/cm2.s Time from removal Dose rate on contact from IRIS sample pit (uSv/hr) (hrs) Cap On Cap Off 0 280 600 1 170 500 35 Dose rate at 10cm from sample (uSv/hr) Cap On Cap Off 35 65 12 22