Download Agilent Technologies 83622A User`s guide
Transcript
[SOURce[1]:]POWer:ALC:SOURce 2 2 3 3 SOURce 1 : POWer:ALC:SOURce 8 9 < INTernal = : DIODe PMETer ; The \[SOURce[1]:]POWer:ALC:SOURce" command selects the type of leveling for output power automatic level control. The parameters are as follows: INTernal Selects internal leveling. DIODe Selects external diode detector leveling. PMETer Selects external power meter leveling. The 4EXT DIODE5 (diode detector leveling), 4EXT METER5 (power meter leveling), and 4INT LEVEL5 (internal leveling) entries in chapter 6 of the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators User's Guide provide detailed information on the type of leveling you select. When the synthesizer is set to the preset state, internal leveling is selected. Query Syntax 2 2 3 3 SOURce 1 : POWer:ALC:SOURce? Returned format: source<NL> Where: source ::= \INT" if internal leveling is currently selected. source ::= \DIOD" if external diode detector leveling is currently selected. source ::= \PMET" if external power meter leveling is currently selected. 2a-7